A coolplate is a process module—not a passive substrate table
After soft bake, post-exposure bake or hard bake, the substrate must reach the defined temperature for exposure, development or metrology. Controlled contact and cooling conditions reduce dependence on uncontrolled waiting time. The real cooling rate still depends on substrate material, dimensions, thickness, starting temperature, target temperature, contact method and the external cooling conditions.
Published CP-200 specifications
| Item | Published specification | Input required for configuration |
|---|---|---|
| Round substrates | Diameter up to 200 mm; thickness up to 14 mm | Material, weight, warpage and backside condition |
| Square substrates | 50 × 50 to 200 × 200 mm | Outline, tolerance and loading method |
| Temperature range | 5–23°C, depending on cooling temperature | Starting temperature, target temperature and allowed variation |
| Installation | Mobile cabinet or bench-top version | Facilities, service clearance and automation requirement |
Where cooling fits in the lithography flow
Cooling after bake
Bring the substrate down to the required temperature before exposure, development or measurement instead of relying on variable ambient waiting time.
Temperature conditioning
Define a target temperature and waiting condition between process steps to support repeatable recipes.
Automated-line takt time
In Duo, ROBO or cluster systems, cooling time affects hotplate utilization, robot scheduling and total system throughput.
Six inputs for coolplate selection
Substrate
Round or square, material, dimensions, thickness, warpage and weight.
Temperature window
Starting temperature, target temperature, allowed variation and stabilization criterion.
Cooling time
State the allowed takt time and validate it with the real substrate rather than an unloaded plate.
Contact and holding
Evaluate proximity or lift pins, substrate-size adaptation and manual vacuum holding.
External cooling
If faster cooling is required, include chiller conditions and facility interfaces in the configuration.
Standalone or integrated
Choose bench-top or mobile cabinet installation, or integration with bake, coat, develop and handling modules.
Configure from the real substrate and temperature curve
Send the substrate drawing, temperature at the end of bake, target temperature, allowed cooling time and automation scope to evaluate the CP-200 configuration.